Users are encouraged to install the latest firmware updates provided by Flow Neuroscience via the Flow app.
Workaround
No workaround given by the vendor.
Thu, 13 Aug 2026 19:30:00 +0000
| Type | Values Removed | Values Added |
|---|---|---|
| Description | An undocumented hard-coded credential, shared by all device units, is authorized to bypass authentication. This allows an attacker within Bluetooth range to arbitrarily manipulate brain stimulation parameters and state. | |
| Title | Flow Neuroscience FL-100 Use of Hard-coded Credentials | |
| Weaknesses | CWE-798 | |
| References |
| |
| Metrics |
cvssV3_1
|
Projects
Sign in to view the affected projects.
Status: PUBLISHED
Assigner: icscert
Published:
Updated: 2026-08-13T19:13:04.494Z
Reserved: 2026-07-28T20:16:40.248Z
Link: CVE-2026-18164
No data.
Status : Received
Published: 2026-08-13T20:17:19.440
Modified: 2026-08-13T20:17:19.440
Link: CVE-2026-18164
No data.
OpenCVE Enrichment
Updated: 2026-08-13T21:15:03Z