Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
Project Subscriptions
No data.
Advisories
No advisories yet.
Fixes
Solution
No solution given by the vendor.
Workaround
No workaround given by the vendor.
References
History
Wed, 03 Jun 2026 02:30:00 +0000
| Type | Values Removed | Values Added |
|---|---|---|
| Description | Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes. | |
| Title | Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface | |
| Weaknesses | CWE-1286 | |
| References |
| |
| Metrics |
cvssV3_1
|
Projects
Sign in to view the affected projects.
Status: PUBLISHED
Assigner: VulnCheck
Published:
Updated: 2026-06-02T19:00:19.249Z
Reserved: 2026-06-02T18:56:43.003Z
Link: CVE-2019-25723
No data.
Status : Received
Published: 2026-06-02T20:16:23.830
Modified: 2026-06-02T20:16:23.830
Link: CVE-2019-25723
No data.
OpenCVE Enrichment
Updated: 2026-06-03T03:30:06Z
Weaknesses